7. What is the maximum pressure the pump can handle?
The pressure depends on the linear force, the diameter of the syringe you are using, and the viscosity.
The linear force of the IPS-12 series (single channel) is 30 kg, IPS-13 series (double dependant channels) is 15 kg per syringe and the IPS-14 serie (independent double channel) series is 30 kg for each syringe.
Please see below some examples of the pressure (calculated using the formula p= F/A)